Dental programs

  Volume size (child mode)
Voxel size, isotropic
Tooth Ø40 x 50 mm
(Ø34 x 42 mm)
Ø40 x 80 mm
(Ø34 x 68 mm)
75 µm*, 100 µm, 150 µm, 200 µm, 400 µm
Teeth Ø80 x 50 mm
(Ø68 x 42 mm)
Ø80 x 80 mm
(Ø68 x 68 mm)
Ø100 x 60 mm
(Ø85 x 50 mm)
Ø100 x 100 mm
(Ø85 x 85 mm)
150 µm, 200 µm, 400 µm
Jaw Ø160 x 60 mm
(Ø160 x 60 mm)
Ø160 x 100 mm
(Ø160 x 100 mm)
Ø200 x 60 mm
(Ø200 x 60 mm)
Ø200 x 100 mm
(Ø200 x 100 mm)
200 µm, 400 µm, 600 µm
Face Ø160 x 160 mm
(Ø160 x 160 mm)
Ø200 x 170 mm
(Ø200 x 170 mm)
200 µm, 400 µm

*Requires Endodontic imaging licence, only for the smallest volume

ENT (Ear, Nose, Throat) programs

  Volume size
(child mode)
Voxel size, isotropic
Nose Ø80 x 80 mm
(Ø68 x 68 mm)
200 µm, 400 µm
Sinus Ø100 x 100 mm
Ø100 x 170 mm
Ø140 x 100 mm
Ø140 x 170 mm
Ø200 x 100 mm
Ø200 x 170 mm
200 µm, 400 µm, 600 µm
Middle ear Ø40 x 50 mm
(Ø34 x 42 mm)
Ø80 x 80 mm
(Ø68 x 68 mm)
75 µm*, 100 µm, 150 µm, 200 µm
Temporal bone Ø80 x 80 mm
(Ø68 x 68 mm)
150 µm, 200 µm
Vertebrae Ø80 x 80 mm
(Ø68 x 68 mm)
200 µm, 400 µm
Airways Ø80 x 80 mm
(Ø68 x 68 mm)
200 µm, 400 µm

*Requires Endodontic imaging licence, only for the smallest volume

Technical data

Anode voltage 54–90 kV
Anode current 1–12 mA
Focal spot 0.5 mm, fixed anode
Image detector Flat panel
Image acquisition 200/360 degree rotation
Scan time 2–33 s
Reconstruction time 2–55 s

Comparison

  3D s 3D Classic 3D Plus 3D Mid 3D Max
3D Dental programs Yes Yes Yes Yes Yes
3D ENT programs - - Yes Yes Yes
ProFace 3D face photo Yes Yes Yes Yes Yes
3D Models scan Yes Yes Yes Yes Yes
Suresmile certification - Yes - Yes Yes
4D jaw motion - - - Yes Yes
2D panoramic imaging Yes Yes Yes Yes Yes
2D cephalometric imaging Yes Yes Yes Yes -

Physical space requirements

  3D Plus or 3D Mid 3D Plus or 3D Mid with cephalostat
Width 118 cm (47 in.) 206 cm (82 in.)
Depth 137 cm (54 in.) 137 cm (54 in.)
Height** 161–239 cm (64–94 in.) 161–239 cm (64–94 in.)
Weight 131 kg (lbs 289) 146 kg (lbs 322)

Minimum operational space requirements

  3D Plus or 3D Mid 3D Plus or 3D Mid with cephalostat
Width 158 cm (63 in.) 225 cm (89 in.)
Depth 175 cm (69 in.) 175 cm (69 in.)
Height** 239 cm (94 in.) 239 cm (94 in.)

**The maximum height of the unit can be adjusted for offices with limited ceiling space.

Dimensions

  3D Plus or 3D Mid
A 1315–2095 mm (51.8–82.5 in.)
B 1610–2390 mm (63.4–94.1 in.)
C 1130 mm (44.6 in.)
D 930 mm (36.6 in.)
E 247 mm (9.7 in.)
F 810 mm (32 in.)
G 1366 mm (53.8 in.)
H 756 mm (29.8 in.)
J Ø1010 mm (39.8 in.)

Example installation

Included in delivery Planmeca ProMax 3D unit with 3D reconstruction server
Minimum set up Client workstation and database server
•    Planmeca Romexis 3D Explorer
•    Database server
•    Planmeca Romexis Image Database
The client workstation and database server can also be in separate computers.
Additional equipment Additional diagnostic workstations with different software configurations
Planmeca Romexis tools:
•    3D Explorer
•    3D Cross Sections module
•    3D TMJ module
•    3D Implant Planning module
•    DICOM module