Tekniset tiedot

Generator Constant potential, microprocessor controlled, operating frequency 66 kHz
X-ray tube Toshiba D-041SB
Focal spot size
0.4 mm according to IEC 60336
Cone diameter 60 mm (2.36 in.)
Rectangular 36 x 45 mm (1.42 x 1.77 in.)
Max. symmetrical radiation field Ø60 mm at SSD 200 mm
Ø60 mm at SSD 300 mm
according to IEC 806
Total filtration min. 2.5 mm Al equivalent at 70 kV
according to IEC 60522
Inherent filtration 1 mm Al equivalent at 70 kV
according to IEC 60522
Anode voltage 60, 63, 66, 70kV
Anode current 8, 7, 6, 5, 4, 3, 2 mA
Exposure times 0.01–2 sec., 24 steps
SSD (Source-Skin Distance) Standard/Long 200 mm (8 in.)/300 mm (12 in.)
Mains voltage 100 V~/110-115 V~/220-240 V~, 50/60 Hz
Duty cycle 1:13.5
Electrical classification Class I Type B
Weight total 29 kg (64 lbs)
tube head with standard cone 4.2 kg (9.3 lbs)
tube head with long cone 4.5 kg (10 lbs)
Colour  White (RAL 9016)

Asennusvaihtoehdot

Standard wall mount

ProX Standard wall mount

Ceiling mount

ProX Remote control panel

Single stud mount

ProX Dental unit mount

Mobile stand

ProX Fixed control panel with double exposure button

Dental unit mount